External control commands, Configure]:external[:control, Configure]:external:loadonin – GW Instek PEL-3000 Series Programming Manual User Manual
Page 67

COMMNAND OVERVIEW
65
External Control Commands
[:CONFigure]:EXTernal[:CONTrol] ................................. 65
[:CONFigure]:EXTernal:LOADonin ................................ 65
[:CONFigure]:EXTernal[:CONTrol]
Set
Query
Description
Configures the unit for external control or disables
external control.
Syntax
[:CONFigure]:EXTernal[:CONTrol] {OFF | VOLTage |
RESistance | RINV }
Related
Commands
[:CONFigure]:EXTernal[:CONTrol]?
Parameter
OFF
Disables external control
VOLTAGE or VOLT
Sets the unit to external
voltage control
RESISTANCE or RES
Sets the unit to external
resistance control
RINV
Sets the unit to external
resistance (inverted)
control
Return Parameter <ASCII string>
Returns the external
control configuration
Example
:EXTernal VOLT
Turns external control voltage control on.
Query example
:EXTernal?
>voltage
The unit uses external voltage control.
[:CONFigure]:EXTernal:LOADonin
Set
Query
Description
The LOADonin settings determine whether the
load is turned on when an external switch is closed
(LOW) or open (HIGH).