GW Instek GDS-800 Series Programming Manual User Manual

Page 28

Advertising
background image

GDS-806/810/820/840 Programming Manual

27

:ACQuire<X>:POINt (query only)

Transfer the displayed waveform data (always 500 points data totally) from the

oscilloscope. Each point is composed by two bytes (the integer value of 16 bits).

The high byte (MSD) will be prior transferred.

Syntax

:ACQuire<X>:POINt

Arguments

<X>

→Specify the channel number (1|2)

Returns

The string of data is following.

#

Data size digit

Data size

Horizontal

scale

Channel indicator Waveform data size

Waveform data

#: Begin a transmission of data string.

Data size digit: Indicate the digits of following data string amount (1 digit).

Data size: the amount of current data string (4 digits).

Horizontal scale: The corresponding

Horizontal scale

of received waveform data (4

bytes). The

Horizontal scale

is indicated by floating point format which compatible

with IEEE 754 standards.

Channel indicator: Show the channel which sent the waveform data.

Waveform data size: The total amounts indicator of waveform data (4 bytes).

Waveform data: The effective waveform data which covers 500 points (1000 bytes)

totally

Example

Transfer the waveform data of channel 1 at 250μs per division:

:ACQuire1:POINt

, the oscilloscope will return the following messages:

Advertising