GW Instek GDS-800 Series Programming Manual User Manual
Page 54

GDS-806/810/820/840 Programming Manual
53
:TEMPlate:POSition:MAX
(Available for firmware 2.0 and above only)
Sets and query the position of maximum template. This command will not change
the position value that stored in flash ROM. Unless you press the “Save & Create”
key in GO/NO-GO edit menu. That means if you exit from the GO/NO-GO
function, we will recall the value from flash ROM.
Before running this command, the oscilloscope has to be initialized as “Normal
Template” mode, please add the command of “:TEMPlate:MODe 0” in order to
change the settings.
Syntax
:TEMPlate:POSition:MAX <NR2>
:TEMPlate:POSition:MAX?
Arguments
<NR2> is the desired template position. The range is from -12.0 ~ +12.0 div.
Returns
<NR2>
Note A: 1 division = 25 dots
Note B: Before input any template editing commands, the oscilloscope has to be initialized,
please add the command of “:GONogo:FUNCtion 1” in order to initial the oscilloscope.