GW Instek GDS-800 Series Programming Manual User Manual
Page 75

GDS-806/810/820/840 Programming Manual
74
STATus
:QUEStionable
:OPERation
Figure 5-2: STATus hierarchy of SCPI defined register
:CONDition
:ENABle
EVENt?
:CONDition
:ENABle
EVENt?
:QUEue
[EVENt]?
Enable
Register
Event
Register
Condition
Register
To SBR
Figure 5-3: Status registers and related commands
The CONDition register is a read-only register which monitors the present state of
the instrument. The CONDition register updates in real time and the inputs are not
latched or buffered. When a condition monitored by the CONDition register
becomes true, the bit for that condition also becomes true (1). When the condition is
false, the bit is 0.
The read-only EVENt register latches any false-to-true change in condition. Once
the bit in the EVENt register is set, it is no longer affected by changes in the
corresponding bit of the CONDition register. The bit remains set until the controller
reads it. The command *CLS (Clear Status) clears the EVENt registers.