Level troll 700 instrument – In-Situ Level TROLL 400/500/700/700H Operators Manual User Manual

Page 10

Advertising
background image

1-800-446-7488

10

www.in-situ.com

Level TROLL 700 Instrument

General

Level TROLL 700

Temperature ranges

1

Operational: -20-80° C (-4-176° F)
Storage: -40-80° C (-40-176° F)
Calibrated: -5-50° C (23-122° F)

Diameter

1.83 cm (0.72 in.)

Length

21.6 cm (8.5 in.)

Weight

197 g (0.43 lb)

Materials

Titanium body; Delrin nose cone

Output options

Modbus/RS485, SDI-12, 4-20 mA

Battery type & life

2

3.6 V lithium; 10 years or 2M readings

External power

8-36 VDC

Memory
Data records

3

Data logs

4.0 MB
260,000
50

Log types

Linear, Fast Linear, Linear Average, Event, Step Linear, True Logarithmic

Fastest logging rate & Modbus rate

4 per second

Fastest SDI-12 & 4-20 mA output rate 1 per second

Real-time clock

Accurate to 1 second/24-hr period

Sensor Type/Material

Piezoresistive; titanium

Calibrated Range
(Usable Depth)

Absolute (non-vented)
30 psia (11 m, 35 ft)
100 psia (60 m, 197 ft)
300 psia (200 m, 658 ft)
500 psia (341 m, 1120 ft)
1000 psia (693 m, 2273 ft)

Gauged (vented)
5 psig (3.5 m, 11.5 ft)
15 psig (11 m, 35 ft)
30 psig (21 m, 69 ft)
100 psig (70 m, 231 ft)
300 psig (210 m, 692 ft)
500 psig (351 m, 1153 ft)

Burst pressure

Max. 2x range; burst > 3x range

Accuracy @ 15° C

4

±0.05% FS

Accuracy (FS)

5

±0.1% FS

Resolution

±0.005% FS or better

Units of measure

Pressure: psi, kPa, bar, mbar, mmHg, inHg, cmH

2

O, inH

2

O

Level: in, ft, mm, cm, m

Temperature Sensor

Accuracy & resolution

±0.1° C; 0.01° C or better

Units of measure

Celsius or Fahrenheit

Warranty

3 years
Up to 5-year extended warranties are available for all instruments—call for details

Footnotes

See See page 8.

Advertising