In-Situ Level TROLL 400/500/700/700H Operators Manual User Manual

Page 78

Advertising
background image

1-800-446-7488

78

www.in-situ.com

Problem

Possible Cause

Possible Solution

Cannot configure level or other
parameters using the Configure
button on the Sensors tab. The
Sensor Setup screen is shown, but
the Configure... button is dim.

The instrument is actively polling
(continually updating real-time readings)
in the Home tab.

The instrument has an active log—a log
that is Ready, Pending, Running, or
Suspended as shown in the Status
column of the Logging tab. Only one
active log can reside in the device at a
time.

Return to the Home tab and stop real-time
readings by clicking the Play button.

Stop or delete the log if possible. Alternatively,
configure parameters after the log is complete.

Advertising
This manual is related to the following products: