Field Deployment

Program the Instrument

In order to set up a log or download data, you must connect the instrument to a
computer running Win-Situ 5 Software, or a RuggedReader Handheld PC running Win-
Situ Mobile. See page 50.

Position the Instrument

Place the instrument at the desired depth. Position the instrument below the lowest
anticipated water level, but not so low that the pressure sensor range might be
exceeded at the highest anticipated water level. Refer to the tables below to determine
usable depth.

A BaroTROLL Instrument can be deployed with a non-vented
instrument to compensate level data for changes in atmospheric
pressure. Make sure the clocks on both instruments are
synchronized, and install the BaroTROLL in a location that will
never be submerged in water.

See page 39.

Range

Effective Range

Usable Depth

PSIA

PSIA

kPA

Meters

Feet

30

15.5

106.9

11

35

100

85.5

589.5

60

197

300

285.5

1968

200

658

500

485.5

3347

341

1120

1000

985.5

6795

693

2273

* Effective range for psia sensors is limited by an estimated
14.5 PSI atmospheric pressure at sea level.