3 how to display fluctuation graph – KANOMAX 6162 Anemomaster User Manual

Page 16

Advertising
background image

2. Basic Operation

10

TIME (20S/DIV)

25

0

hold

VEL. SLOW2

2.3 How to Display Fluctuation Graph

Š You can monitor velocity fluctuation for 90 seconds.

(* This function cannot be used in the calculation mode.)

Display

Key Procedure

▽

Press ▽ key.

▽

▽

FAST/SLOW key is useful when …

>>> reading varies significantly and it is hard to see the graph.

Press FAST/SLOW key.

····· SLOW 1 (moving average deviations for 5 sec.)

FAST/SLOW

FAST/SLOW Press FAST/SLOW key

····· SLOW 2 (moving average deviations for 10 sec.)

△

Press △ key to go back to the original screen.

HOLD

PRINT

Press HOLD key

★

Display screen will be on hold.

Even while the reading is on hold, you can still change the

range and switch FAST/SLOW.

Press PRINT key

★

When the reading is on hold, the display can be printed out.

08.07.01 FAST

10:26:40

5.38 m/s

25.7 ℃

TIME (20S/DIV)

m/s VEL. FAST

50

0

TIME (20S/DIV)

m/s VEL. FAST

25

0

TIME (20S/DIV)

m/s VEL. SLOW1

25

0

TIME (20S/DIV)

m/s VEL. SLOW2

25

0

08.07.01 SLOW2

10:26:40

5.38 m/s

25.7 ℃

Air velocity value

updated every 1 sec
one screen: 90 sec

air

velocit

y

measuring time

(sec)

m/s VEL. FAST

TIME (20S/DIV)

50

0

▽

key

Range: 50, 25, 10, 5, 2, 1 m/s

FAST, SLOW1, SLOW2

"

P9

FAST/SLOW key

20 sec

90 sec

time axis

Press ▽ key again.

★

Air velocity range changes.