Main specification – KANOMAX 6162 Anemomaster User Manual

Page 42

Advertising
background image

6. Main Specification

36

6. Main Specification

Specification

(1) Model Name

Main Unit:

Probe:

Model 6162

Model 0203 (for mid temperature)

Model 0204 (for high temperature)

Model 0205 (for high temperature)

(2) Measurement Feature

Air velocity and air temperature (simultaneous measurement)

(3) Measurement Object

Clean air under normal pressure and normal temperature

Air Velocity: V

0

~

50.0m/s

(4) Measurement Range

Air Temperature:

0

~

200 °C (MODEL 0203)

0

~

500 °C (MODEL 0204)

0

~

500 °C (MODEL 0205)

(5) Measurement Accuracy

(Normal Temp. Accuracy:

18~28 °C)

(6) Temperature

Compensation Accuracy

(Air Velocity)

(7) Heat Resistance of Cable

Teflon Coated (probe side): 200 °C

Vinyl Code (connection cable): 80 °C

(8) Response Air

Velocity:

Air Temperature:

4 sec (90% response at air velocity of 5 m/s)

5 sec (90% response at air velocity of 5 m/s)

(9) Display Screen

Graphic LCD (120×64dot):

displaying velocity and temperature simultaneously

w/t backlight and brightness adjustment functions

(10) Memory Capacity

Max 999 data (when measuring in one page)

(11) Input & Output Terminal

Remote Terminal:

Analog output Terminal:

Digital Output Terminal:

START/STOP function

simultaneous velocity & temp. output

output voltage 0

~

1V (Output Impedance: 47Ω)

(Output Accuracy: 0.5%FS)

RS-232C (serial)

0 ~

99 °C : V

0

= 0.2 m/s

100 ~ 199 °C : V

0

= 0.4 m/s

200 ~ 299 °C : V

0

= 0.7 m/s

300 ~ 399 °C : V

0

= 1.0 m/s

Measurement Range

Specified Accuracy Display Resolution

V

0

~ 4.99 m/s

5.00 ~ 9.99 m/s

0.01 m/s

10.0 ~ 24.9 m/s

Air

Velocity

25.0 ~ 50.0 m/s

±3%FS

0.1 m/s

0 ~ 99.9 °C

0.2 °C

100 ~ 199 °C

Air

Temp.

200 ~ 500 °C (0204/0205)

±1%±1˚C of reading

1 °C

MODEL 0203

MODEL 0204/0205

Temp.

Velocity

0 ~ 200 °C

0 ~ 400 °C

V

0

~ 4.99 m/s

±10%FS

±15%FS

5.00 ~ 9.99 m/s
10.0 ~ 24.9 m/s
25.0 ~ 50.0 m/s

±6%FS ±10%FS

Advertising