Altosonic – KROHNE UFM 600T EN User Manual
Page 31
C.2
Start Up Menu
C.2.1
Enter start up menu
The Start Up Menu is activated by pressing the key twice
from the measuring mode or pressing the key once from
the main menu 1.0.0.
C.2.2
Application Functions
The functions in which the application parameters can be
programmed always appear on the display in the same order
as described below:
To enter a next function, the key must be pressed.
DIAMETER
Program the outer pipe diameter.
The diameter can be entered either in mm or in inches.
Range:
2.0000 E 0 mm
≤ DIAMETER ≤ 4.0000 E 3 mm
7.8740 E-2 inch
≤ DIAMETER ≤ 1.5748 E 2 inch
PIPEWALL
Program the pipe wall thickness.
The wall thickness can be entered either in mm or in inches.
Range:
1.0000 E-1 mm
≤ PIPEWALL ≤ 5.0000 E 1 mm
3.9370 E-3 inch
≤ PIPEWALL ≤ 1.9685 E 0 inch
30
ALTOSONIC
Operation
UFM 600 T