KROHNE UFM 600T EN User Manual
Page 70

3.5.4 FACT. QUANT.
If a user defined unit is applied the quantity factor has to be
programmed.
This must be the conversion ratio into m3.
In the example with Barrels/Day (see 3.5.3) the quantity unit
Barrel has to be converted into m3 (1 Barrel = 1.5898 E-1
m3).
3.5.5 FACT. TIME
If a user defined unit is applied the time factor has to be pro-
grammed.
This must be the conversion ratio into seconds.
In the example with Barrels/Day (see 3.5.3) the time unit Day
has to be converted into seconds (1 Day = 8.6400 E 4 sec-
onds).
3.5.6 SET DATE
Program the correct date in this function.
The actual date is included in the header of each data block
that is stored.
On entering this function the date can be programmed in
(mm-dd-yy) notation.
Programming the actual date is similar to that of numbers.
3.5.7 SET TIME
Program the correct time in this function.
The actual time is included in the header of each data block
that is stored.
Programming is similar to numbers.
67