C.3.2
Error messages
If the microprocessor produces an error message a * sign
appears at the lower left hand side of the display.
Enter the QUIT / RESET mode to view the error message(s)
(see chapter C.5).
It depends on the settings of the Installation functions 3.2.4
and 3.2.6, whether the error messages will be displayed in
between the flow data.
List of Error Messages:
SIGN LOST
:
Transducer disconnected
TOTALIZER
:
Overflow of the totalizer.
FREQ.OUTP. F:
Overflow of the frequency signal; full
scale value is too small
CUR. OUTP. I :
Overflow of the current signal; full
scale value is too small
CAL. DATA
:
The transducer calibration is
distorted; contact your supplier
EE1 EE2
:
Error detected during memory check;
switch unit off and on; if message still
reappears after two attempts, contact
your supplier
EEPROM2
:
See EE1 EE2
ROM
:
Checksum error detected during
memory test; contact your supplier
RAM
:
Error detected during verification of
parameter settings stored previously;
use function 3.5.8 DEFAULT SET and
switch off and on; if message
reappears contact your supplier.