D.6
Parameter errors
If discrepancies in the parameters entered are found the
microprocessor will display the message "PARAMERROR"
thus preventing you from storing parameters that are incor-
rect.
The PARAMERROR menu can appear at the end of both the
Start Up menu and the Installation Menu.
A third main menu is available now:
Fct. 4.0.0 PARAM.ERROR
In this menu only one submenu is available at a time; only the
submenu that applies to the discrepancy to be corrected will
be displayed.
List of possible submenus and functions within PARAMER-
ROR:
4.0.0
PARAMERROR
Discrepancy:
4.1.0
V RANGE
The combination of the diame-
ter, wall and lining thickness and the full
scale flow rate is not feasible.
4.1.1
DIAMETER
4.1.2
PIPEWALL
4.1.3
LINING TH.
4.1.4
FULL SCALE
4.2.0
I RANGE
[I 100 PCT] does not exceed
[I 0 PCT] by at least 4 mA
4.2.1
I 0 PCT
4.2.2
I 100 PCT
4.3.0
I MAX mA
[I MAX] is not equal to or greater
than [I 100 PCT]
4.3.1
I 100 PCT
4.3.2
I MAX
4.4.0
CUTOFF
CUTOFF ON exceeds CUTOFF OFF
74
ALTOSONIC
Programming mode