Operation – KROHNE MFC 400 EN User Manual
Page 114

6
OPERATION
114
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
Typical frequency values (at 20°C / 68°F; ±5 Hz)
Problems with zero calibration
1 Stop the flow.
2 Set the totaliser in Fct. C3.y.1 to "Absolute Total".
3 Set the low flow cutoff in Fct. C3.y.3 to zero.
4 Perform automatic zero calibration.
5 Reset totaliser and count for 2 minutes.
6 Compare the added value to the specified zero calibration stability.
Possible causes of poor zero calibration:
• Valves not completely closed, air or gas bubbles or deposits on measuring tube
Size
All materials
Empty
Water
6000 - 08
220
205
6000 - 10
187
172
6000 - 15
212
191
6000 - 25
266
231
6000 - 50
262
230
6000 - 80
234
203
6000 - 100
257
221
6000 - 150
245
208
6000 - 200
218
184
6000 - 250
187
160
INFORMATION!
For best results, perform zero calibration with the medium at operating temperature.
.book Page 114 Monday, August 26, 2013 9:05 AM