4.4.1 Connection of signal cable - Field housing and connection box for sensor
1 Unscrew the terminal compartment cover.
2 Pass the prepared signal cable through the cable entry.
3 Secure the signal cable using the clip.
4 Connect the electrical conductors as shown. The shielding must also be connected to the
spring terminal.
5 Re-fit the cover and tighten it by hand.
Figure 4-1: Connection of signal cable - Field housing and connection box for sensor
Cable
Cable
Connection terminal
Cable pair
Colour
1
yellow
X1 SA+
1
black
X1 SA-
2
green
X1 SB+
2
black
X1 SB-
3
blue
X2 T1
3
black
X2 T2
4
red
X2 T3
4
black
X2 T4
5
white
X3 DR+
5
black
X3 DR-
INFORMATION!
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin-
free and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.