TECHNICAL DATA
8
125
MFC 400
www.krohne.com
08/2013 - 4002075802 - MA MFC 400 R02 en
Materials
Signal converter housing
Standard: die-cast aluminium (polyurethane coated)
Option: stainless steel 316 (1.4408)
Measuring sensor
For housing material, process connections, measuring tubes, accessories and
gaskets, refer to technical data for the measuring sensor.
Electrical connection
General
Electrical connection is carried out in conformity with the VDE 0100 directive
"Regulations for electrical power installations with line voltages up to 1000 V" or
equivalent national specifications.
Power supply
Standard: 100…230 VAC (-15% / +10%), 50/60 Hz
Option 1: 24 VDC (-55% / +30%)
Option 2: 24 VAC/DC (AC: -15% / +10%, 50/60 Hz; DC: -25% / +30%)
Power consumption
AC: 22 VA
DC: 12 W
Signal cable
Only for remote versions.
10 core shielded cable. Detailed specifications are available on request.
Length: max. 20 m / 65.6 ft
Cable entries
Standard: M20 x 1.5 (8...12 mm)
Option: ½ NPT, PF ½
.book Page 125 Monday, August 26, 2013 9:05 AM