Criteria of judgment and cause

10-2. Criteria of judgment and cause

Check

Criteria of Judgment

Cause

Frequency

Frequency of V1 is within 45 - 65Hz.

- Voltage clip is firmly connected to the

DUT?

- M e a s u r i n g t o o h i g h h a r m o n i c

components?

Voltage input

Voltage input is 10% or more of
(Voltage Range x VT).

- Voltage clip is firmly connected to the

DUT?

- Voltage test leads are firmly connected

to the Voltage input terminal on the
instrument?

Voltage balance

Voltage input is within ±20% of reference

voltage (V1)
* (not checked in single-phase wiring)

- Settings are matched with the wiring

system under test?

- Voltage clip is firmly connected to the

DUT?

- Voltage test leads are firmly connected

to the Voltage input terminals on the
instrument?

Voltage phase

Phase of voltage input is within ±10º of
reference value (proper vector).

- Voltage test leads are properly

connected?

(Connected to proper channels?)

Current input

Current input is 10% or more and 110%
or less of (Current Range x CT).

- Clamp sensors are firmly connected

to the Power input terminals on the
instrument?

- Setting for Current Range is appropriate

for input levels?

Current phase

- PFi (absolute value) is 0.5 or more.
(3P3W3A : 0≦PFi)
- Pi is positive value.

- Arrow mark on the Clamp sensor

and the orientation of flowing current
coincide with each other?

(Power supply to Load)
- Clamp sensors are connected properly?