Warning, Caution, 5 measuring pressure – Martel Electronics MC1010 User Manual

Page 15: O, cmh

Advertising
background image

15

3.5 Measuring Pressure

Note: The MC1010 is compatible with BETA Calibrator Pressure Modules. The accessory
BPPA-100 is required for pressure measurement.

Note: Pressure is not read from modules with frequency or pulse train mode enabled.

Note: On high pressure modules engineering units normally associated with low pressure
ranges such as, inH

2

O, cmH

2

O, etc. are not valid selections. Selecting one of this units with a

high pressure module attached will cause the display to read "----".

Warning!

To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly
bleed off the pressure before you attach the pressure module to the pressure line.

Caution

To avoid mechanically damaging the pressure module, never apply more than 10 ft-lb. of
torque between the pressure module fittings, or between the fittings an the body of the module.

To avoid damaging the pressure module from overpressure, never apply pressure above the
rated maximum printed on the module.

To avoid damaging the pressure module from corrosion, use it only with specified materials.
Refer to the pressure module documentation for material compatibility.

To measure pressure, follow these steps:

1.

Connect the pressure module to the calibrator as shown in Figure 10.

Note: Make sure the calibrator is on before you plug in the pressure module.

2.

Select the [CONFIG] option from the main menu.

3.

Select [PRESSURE] from the primary parameters.

4.

Select the desired measuring unit.

5.

Zero the pressure module. The zero function on the calibrator can be found in the
pressure zeroing menu.

Figure 10. Connections for Measuring Pressure

PRESSURE

MODULE

VALVE

F1

F2

F3

CE

ENTER

8

9

4

1

5

2

0

6

3

.

7

HOME

Advertising