4.7 Sourcing a pulse train

The calibrator can produce a pulse train with an adjustable number of pulses at a desired
frequency. For example, setting the frequency to 60Hz and the number of pulses to 60 would
produce 60 pulses for a period of 1 second. To source a pulse, use the same connection as
for frequency, and proceed as follows:

1.

Select the [CONFIG] option from the main menu and select pulse from the primary
parameters.

2.

Choose the desired unit and enter the frequency using the keypad.

3.

Select the [COUNTS] function from the home menu to enter the number of pulses. Use
[TRIG] to start and stop the signal.

4.

The amplitude of the pulse can be adjusted in the same manner as for frequency.

Figure 15. Connections for Outputting Thermocouples

4.8 Sourcing Thermocouples

To source a thermocouple use the following steps:

1.

Connect the thermocouple leads to the appropriate polarized TC miniplug, and insert the
plug into the TC terminals on the calibrator, as shown in Figure 15.

2.

Select the [CONFIG] option from the main menu, and choose thermocouple [TC] from
the primary parameters.

3.

Choose output [OUT] from the input/output control.

4.

Select the desired thermocouple type from the sensor types.

5.

Enter the temperature using the keypad.

MEASURE / SOURCE

3W
mA+

TC

+

V

Hz

Ω

4W
mA–

+

+

TC PLUG

Note: TC wire used must match
the thermocouple type being calibrated.

PULSE

5.0 Vpp

MENU

1.0 kHz

TRIG

COUNTS

FREQ OUT

5.0 Vpp

MENU

1.0 kHz

LIGHT

STEP