Martel Electronics MC1010 User Manual

Page 36

Advertising
background image

36

FREQ_UNIT?

Returns the frequency unit currently being used by the frequency and pulse modes.

FUNC?

Returns the current mode of the display. For example if the calibrator is set to volts FUNC?
Would return:

DCV

LOCAL

Restores the calibrator to local operation if it was in remote mode. Also clears LOCKOUT if the
unit was in lockout mode.

MEAS

Sets the display to measure mode. The command is followed by any of the parameters
except for pulse and mA sim, which are source only. Enter DCI for mA, DCV for volts, TC for
thermocouple, RTD for RTD, FREQUENCY for frequency, and PRESSURE for pressure. For
example:

MEAS DCV sets the display mode to VOLTS IN

PRES_UNIT

Sets the unit for measuring pressure on the display. Add the unit after the command. The
available pressure units and their syntax are shown in the Table 7. (Parameter Units).
For example:

PRES_UNIT KPAL sets the pressure unit to kiloPascals

OUT

Sets the output of the calibrator. This command may be used to output mA, volts, frequency,
temperature, and ohms. Frequency output, which is set by the command FREQ_TYPE, is
either continuous or pulse. The calibrator is automatically set to source mode when OUT is
entered. A number and its unit must follow the command. See Table 7. (Parameter Units) for a
list of available units. For example:

OUT 10 MA sets the calibrator to mA OUT mode and sets the output to 10mA.

OUT?

Returns the output of the calibrator. Using the above example, OUT? Would return:

1.000000E-02, A

PRES?

Returns the model and serial number of the attached pressure unit. Returns NONE if no
pressure unit is attached. For example:

PRES? Will return MARTEL,001PNS,3,0

Advertising