Martel Electronics MC1010 User Manual

Page 28

Advertising
background image

28

Table 5: Common Commands

Command

Description

*CLS

*CLS (Clear status.) Clears the ESR, the error queue, and the RQS bit in the status byte.
Terminates pending Operation Complete commands

*ESE

Loads a byte into the Event Status Enable register.

*ESE?

Returns the contents of the Event Status Enable register.

*ESR?

Returns the contents of the Event Status register and clears the register.

*IDN?

Identification query. Returns the manufacturer, model number, and firmware revision level
of the Calibrator.

*OPC

Enables setting of bit 0 (OPC for "Operation Complete") in the Event Status Register to 1
when all pending device operations are complete.

*OPC?

Returns a 1 after all pending operations are complete. This command causes program
execution to pause until all operations are complete.

*RST

Resets the state of the instrument to the power-up state. This command holds off
execution of subsequent commands until it is complete.

*SRE

Loads a byte into the Service Request Enable register.

*SRE?

Returns the byte from the Service Request Enable register.

*STB?

Returns the status byte.

*WAI

Prevents further remote commands from being executed until all previous remote
commands have been executed.

Table 6: Calibrator Commands

Command

Description

CAL_START

Puts the calibrator in calibration mode

CJC_STATE

Turns CJC on or off.

CJC_STATE?

Returns the state of the CJC

CPRT_COEFA

Sets the custom RTD coefficient A

CPRT_COEFA?

Returns the custom RTD coefficient A

CPRT_COEFB

Sets the custom RTD coefficient B

CPRT_COEFB?

Returns the custom RTD coefficient B

CPRT_COEFC

Sets the custom RTD coefficient C

CPRT_COEFC?

Returns the custom RTD coefficient C

CPRT_MIN_T

Sets the custom RTD minimum temperature

CPRT_MIN_T?

Returns the custom RTD minimum temperature

CPRT_MAX_T

Sets the custom RTD maximum temperature

CPRT_MAX_T?

Returns the custom RTD maximum temperature

CPRT_R0

Sets the custom RTD R0 resistance

CPRT_R0?

Returns the custom RTD R0 resistance

FAULT?

Returns the error code of an error that has occurred

FREQ_LEVEL

Sets the frequency and pulse amplitude

FREQ_LEVEL?

Returns the frequency and pulse amplitude

Advertising