7. Maintenance / Warranty

7.1 Replacing Batteries

Replace batteries as soon as the battery indicator turns on to avoid false measurements. If
the batteries discharge too deeply the MC1010 will automatically shut down to avoid battery
leakage.

Note: Use only AA size alkaline batteries or optional rechargeable batteries.

Warning

This product has non-rechargeable alkaline batteries installed.

Prior to connecting any AC adapter or battery charger, these batteries must be removed
and/or replaced with rechargeable batteries such as Nickel Cadmium or Nickel Metal Hydride
types.

Failure to remove the batteries will result in damage to the calibrator and voids the warranty.

7.2 Cleaning the Unit

Warning

To avoid personal injury or damage to the calibrator, use only the specified replacement parts
and do not allow water into the case.

Caution

To avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers.

Clean the calibrator with a soft cloth dampened with water or water and mild soap.

RTD Type

Range (°C)

Accuracy (°C)

Minimum

Maximum

Measure 4W/Source

Measure 3W

100.0

260.0

0.4

0.4

260.0

300.0

0.4

0.4

300.0

400.0

0.4

0.5

400.0

630.0

0.5

0.5

PT385, 1000 ohm

-200.0

-80.0

0.2

0.2

-80.0

0.0

0.2

0.2

0.0

100.0

0.2

0.2

100.0

260.0

0.2

0.2

260.0

300.0

0.3

0.3

300.0

400.0

0.3

0.3

400.0

630.0

0.4

0.4

NI120

-80.0

260.0

0.1

0.2

Cu10

-100.0

260.0

1.3

2.5

Cu50

-180.0

200.0

0.3

0.5

Cu100

-180.0

200.0

0.1

0.3

YSI 400

15.0

50.0

0.1

0.1