5 MEASUREMENT

5.1 Display in Measurement mode

When the controller is initially powered on, it automatically enters into the

Measurement mode after the large dual LCD displays all segments briefly.

The upper display shows the measured pH or ORP value, while lower display

shows either temperature value if controller is set for pH measurement or “OrP”

if it is set for ORP measurement.

Annunciators at right side of display indicate whether controller is set for pH or

mV measurement.

Similarly annunciators or icons at top or left side of display shows current status

of controller, e.g. “MEAS”, “SETUP”, “CAL”, “READY”, etc. In addition, error

messages also guide user in time of any faulty conditions as indicated by “ERR”

or icons flashing.

5.1.1 Check

electrode

performance

To read current electrode slope and offset values:

1) Press CAL key followed by ENTER key. The upper display shows

electrode slope. The lower display reading shows the pH reading at 0 mV.

2)

Press

Δ and keys to return to Measurement mode.