6.2

ORP – mV Calibration

This mode allows one-point calibration.

1) Enter Calibration mode. While in Measurement

mode, push CAL key. Press ENTER key. The upper

and lower display reads “CAL OrP”.

Note: If upper and lower display read “CAL PH”, see

section 7.4 for procedures on how to switch from pH

to ORP mV readings.

2) Place sensor in ORP solution.

3) Press ENTER key to begin calibration. “CAL”

indicator appears on display. Upper display shows

current mV output of electrode without any offset

adjustment.

4) Press

Δ or keys to adjust mV value to that of standard solution.

5) To confirm, press ENTER key. The controller calculates the correction factor

and returns to Measurement mode.

CAL

OrP

SETUP