7.3.4

Selecting a hysteresis (dead band) value


Hysteresis prevents rapid contact switching if your value is fluctuating near set

point. It does this by overshooting the set point value to a specified hysteresis

value. You can set the hysteresis value from 0.1 to 1.0 pH, 10 to 100mV.

Example: You have set your high set point at pH 9.0 and your hysteresis limit

value at pH 0.5. If your measured value overshoots high set point pH 9.0, the

controller’s relay activates, which in turn activates external devices such as a

pump or valve. The actions of the external device will cause the solution’s pH to

drop. When the pH drops to 8.5 pH, the relay and hence the pump will switch off.

1) Follow directions in 7.3.1 to enter Control

Relay mode.

2) Press ENTER key. Scroll with

Δ or keys

until upper display shows the hysteresis (dead

band) value and lower display shows “HYS”.

3) Press

Δ or keys to enter your hysteresis value for Set point 1 (Set point

2). Your controller will activate at the value you select.

4) Press ENTER key to confirm your selection.

5) Proceed to 7.3.5 step 3, or return to Measurement mode by pressing

Δ and

keys simultaneously (escape).

NOTE: Please refer to Appendix 3 for a graphical representation of the

Hysteresis.