2 setting temperature (set, C) sub-function – Thermo Fisher Scientific Alpha pH 800 User Manual
Page 24

17
Note: Offset value is reset during full calibration. See section 6 for full calibration
instructions.
Note: The offset parameter is blanked out during ORP operation.
7.2 Setting temperature (Set
o
C) sub-function
7.2.1
Selecting automatic or manual temperature compensation
1) Enter Advanced Set-up mode by pressing
ENTER key.
2) Press
Δ or ∇ keys to scroll until display
shows “Set
o
C”.
3) Press ENTER key. The lower display shows
“Atc”; upper display shows “on” or “oFF”
depending on whether or not ATC is selected.
4) Press
Δ or ∇ keys to toggle between ATC
on and off.
5) Proceed with additional Advanced Set-up
procedures (press ENTER again) or return to
Measurement mode by pressing
Δ and ∇ keys
(escape) simultaneously.
oFF
Atc
on
Atc
SETUP
HOLD
SETUP
SEt
o
C
SETUP
HOLD
HOLD