2 setting temperature (set, C) sub-function – Thermo Fisher Scientific Alpha pH 800 User Manual

Page 24

Advertising
background image

17

Note: Offset value is reset during full calibration. See section 6 for full calibration

instructions.

Note: The offset parameter is blanked out during ORP operation.

7.2 Setting temperature (Set

o

C) sub-function


7.2.1

Selecting automatic or manual temperature compensation

1) Enter Advanced Set-up mode by pressing

ENTER key.

2) Press

Δ or keys to scroll until display

shows “Set

o

C”.

3) Press ENTER key. The lower display shows

“Atc”; upper display shows “on” or “oFF”

depending on whether or not ATC is selected.

4) Press

Δ or keys to toggle between ATC

on and off.

5) Proceed with additional Advanced Set-up

procedures (press ENTER again) or return to

Measurement mode by pressing

Δ and keys

(escape) simultaneously.

oFF

Atc

on

Atc

SETUP

HOLD

SETUP

SEt

o

C

SETUP

HOLD

HOLD