Er 2 - External RAM

1. Turn off the power to the instrument. Wait 5 seconds,

Failure

and turn the power on.

Er 3 - EPROM Checksum

1. Perform the steps listed for Er 1 except that

Failure

the EPROM (U7) on the Processor board should be
inspected.

Er 4 - RTD Mismatch Error

1. Check the connections to the instrument for the RTD
Input Calibration (CAL5) as described in the
Calibration section (page 37). Repeat the RTD Input
Calibration.

Er 5 - No Zero Crossings

1. Turn off the power to the instrument. Wait 5 seconds,

Detected

and turn the power on.
2. Connect the instrument to another AC power source.

Er 6 - AC line below 43 HZ

1. Turn off the power to the instrument. Wait 5 seconds,
and turn the power on.
2. Connect the instrument to another AC power source.

Er 7 - AC line over 64 HZ

1. Turn off the power to the instrument. Wait 5 seconds,
and turn the power on.
2. Connect the instrument to another AC power source.

Er 8 - Cal 2 Volt Input

1. Check that 50 mVDC is properly connected to the

Error

instrument and is within the tolerance limits as
indicated in the CAL2 procedure of the Calibration
section (page 35).
2. Turn off the power to the instrument and open the
cover. Loosen the platen hold down screw and swing
the platen open. Inspect the Processor board
(Appendix A-1, page 50) to insure that the input
conditioning jumper JU4 is in the non-volt position.
The jumper must be in the non-volt position to perform
the CAL 2 procedure properly. Close the platen
and tighten the hold down screw. Close the cover and
turn on the power to the instrument.
3. Perform the CAL2 procedure as described in the
Calibration section (page 35).

Er 9 - ADC Reference

1. Perform the CAL2 procedure as described in the

Number Error

Calibration section (page 35).

Er10 - ADC Reference

1. Perform the CAL2 procedure as described in the

Voltage Error

Calibration section (page 35).

Er11 - Cold Junction

1. Perform the CAL3 procedure as described in the

Compensation Error

Calibration section (page 36).