Troubleshooting – Badger Meter Fluid Management System User Manual

Page 24

Advertising
background image

Electronic Preset Meter Model EPM-3, RF 2.4 GHz

TROUBLESHOOTING

Pressurized Equipment

Relieve the pressure prior to checking or repairing the meter. Make sure all valves, controls and pumps are operating correctly.

Symptom

Possible Cause

Remedy

Battery Icon is

displayed

Batteries are low

Replace batteries

Display Blank

Meter asleep

Press RESET button

Loose battery connection

• Remove battery pack and check battery connection
• Press RESET button

Batteries dead

• Replace batteries
• Press RESET button

Meter does not latch

for batching

Meter not in AUTO mode

Press AUTO button and program batch size

Meter not reset after prior batch

Press RESET button

Low batteries

• Check for battery icon
• Replace batteries
• Press RESET button

Slow or no fluid flow

Filter is clogged

Clean or replace the filter in the swivel nut

Pump pressure is low

Turn up the pump pressure

Foreign material is jamming meter Contact your local distributor for repair

Meter inaccurate

Scale factor not correct for fluid

Enter program mode, check and reset program factor

Page 24

November 2012