MN179-1 _ 02/14 AirSep Corp. - FreeStyle 5 Service Manual
Rev B
4. If either of the two fittings connecting the compressor to the manifold is connected to the
compressor, remove it and insert it into the manifold. (See Figure 5.20.)
Figure 5.20
5.
Connect the new compressor to the PSA assembly, and tighten the bolt to a torque of 15 in-
lbs. (See Figure 5.21.) Be sure the gasket is in place.
Figure 5.21
5.5.2 Compressor Assembly Installation
To install the new compressor assembly, perform the compressor removal
procedure in reverse order.