Problem, Possible cause, Solution – AirSep MN179-1 B User Manual

Page 55

Advertising
background image

MN179-1 _ 02/14 AirSep Corp. - FreeStyle 5 Service Manual

Rev B

Problem

Possible Cause

Solution

Battery duration is less

than per specification.

Note: While charging,
battery may not initiate a
charge. Allow unit battery
time to drop to operating
temperature before charging.
Will initiate in approx. (20
min).

Note: Whenever storing
FreeStyle 5 for an extended
period of time, ensure that
the battery the state of
charge (SOC) is not depleted
or stored at 100%. Both
these conditions can result in
unrecoverable capacity loss.
The recommend SOC is
25% to 50% for long-term
storage. The recommended
storage temperature is 23º C
± 2º C.

Battery not fully charged.

Battery has been stored in
a discharged or 100%
charged state for
prolonged period of time or
stored in a very high
temperature environment.

Battery reached cycle life.

Ensure that battery
charges for at least

3½ hours and all (4) LEDs

are lit solid.

Replace battery.


Replace battery.

Unit does not
start on battery.

Unit temperature has
exceeded battery operating
temperature

5 to 40ºC.

Battery reached cycle life
Battery is defective

Allow unit to cool

before starting.



Replace battery.

Battery does not charge.

Unit temperature has
exceeded battery operating
temperature

5 to 40ºC.

Battery is defective.

AC adapter connector is
defective or contact pin is
bent.



Defective AC adapter

Allow battery to cool before

charging.


Replace battery

Replace battery.

Replace AC adapter.