7 solenoid valve replacement – AirSep MN179-1 B User Manual

Page 37

Advertising
background image

MN179-1 _ 02/14 AirSep Corp. - FreeStyle 5 Service Manual

Rev B

5.7 Solenoid Valve Replacement

1. Remove back cover panel per Section 5.2.2.

2. Determine which valves need to be replaced.

3. Remove the tie-wraps that bundle the value wires.

4. Carefully disconnect the Molex connector from the circuit board by pulling on the
Molex connector, not the valve wire.

5. Using a small Phillips screwdriver, remove the two screws fastening the valve.

6. Remove the valve.


7.

Verify two o-ring seals are installed on the new valve.

8. Fasten new valve using the two screws provided.

9. Re-bundle wires and tie-wrap.

10. Start FreeStyle 5 unit and leak test around the area where the

o-rings are located on the valve.

It may not be necessary to completely remove the PSA system assembly to replace the
solenoid valve. The Feed B/Waste B OCD valve will require the PSA system to be partially
removed.

Advertising