Sitter Elite

Setting Alarm Mode

(Continued)

4. LED light blinks green to indicate selected mode is in

use (fig. 6). LED blinks yellow if the alarm mode is set to

“Mute” (fig. 6).

5. If alarm is set to “Voice Only” or “Voice and Tone” mode

and custom recorded message does not play, re-record

message. If this does not work, remove batteries for at

least 1 hour to reset alarm to factory default. You can

then re-record the message.

The last option selected is the mode utilized when the

sensor is activated or magnet is removed from the face

plate, depending on which monitoring mode is selected.
6. If you are using the nurse call cable, check to make sure

it is plugged into the correct wall jack and the alarm.

Make sure the mode you select alerts at the nurse’s

station when the alarm is activated.

7. When “Voice Only” or “Mute” are selected, you must have the nurse call cable plugged

into your alarm and the appropriate wall jack for your nurse call system. Ensure that the

nurse call system indicators (nursing station and wall light) are activated when the cable is

unplugged from the wall jack.

8. Replace the programming control protective cover, securing it into place.
To test the modes requiring nurse call interface, remove pressure from the sensor, unfasten the

chair belt sensor, or remove the magnet from the face plate. Check at the nurse’s station to ensure

that the indicator for the appropriate patient alarms on the nurse call system and the wall light,

indicating there is no pressure on the sensor, chair belt sensor is unfastened, or the magnet has

been removed from the face plate.

H O L D / S U S P E N D

®

Sitter

MODE

LED

Fig. 6