Monitoring with a sensor, Sitter elite 23 – Posey Sitter Elite® User Manual

Page 23

Advertising
background image



Sitter Elite

23

Over Mattress Chair Pad, Chair Belt and Toilet Sensors

OVER MATTRESS AND CHAIR SENSOR PAD

In addition to the appropriate steps above:

1. Test several places along the entire

surface of the sensor by applying and

removing pressure to make sure the

alarm activates when pressure is removed

from the sensor/mattress (figs. 29a, b) or

when you unfasten the chair belt sensor

(fig. 31). If the alarm and/or sensor do

not function properly, remove the alarm

and sensor from service and replace them with a properly functioning alarm and/

or sensor. DO NOT use the alarm or sensor if it does not activate each time weight is

removed from the sensor, or the chair belt sensor is unfastened.

2. Make sure sensor pad air intake (“neck” of over mattress or

chair sensor pad) is clear and not blocked (fig. 30). Air must

flow freely in and out of sensor for alarm to function. Make

sure liquid does not enter at “neck” of sensor pad, as this

will damage sensor. If needed, use an incontinence pad to

protect sensor from urine or other liquids.

TESTING CHAIR

SENSOR PAD

Fig. 29b

TESTING OVER

MATTRESS

SENSOR PAD

Fig. 29a

Monitoring with a Sensor

(Continued)

FOR SAFE USE IN ALL SENSOR MODES:

To reduce the risk of serious injury or death, ALWAYS follow these steps after putting the

sensor in place and before leaving patient unattended (see instructions below). DO NOT

use any alarm or sensor that does not alarm each time it is tested.
1. Remove and store magnet when a sensor is in use. If magnet is on the alarm face plate,

the alarm will NOT activate if sensor is disconnected. Store magnet in a safe place, out of

the patient’s reach.

2. Make sure alarm is ON and in monitoring mode (LED Mode light flashing green or

yellow (“Mute” mode only).

3. Check that the RJ11 plug on the sensor cable is not damaged (plug broken, or wires

disconnected) and is securely connected to the alarm.

4. Disconnecting the sensor from the alarm will cause the alarm to activate. This is called

a “failsafe” mode. Disconnect the sensor to make sure the failsafe mode works. DO

NOT use the alarm if the alarm does not sound when the sensor is disconnected.

5. When connecting the alarm to the nurse call system, check that the nurse call cable is

securely connected to the alarm and the nurse call panel. ALWAYS test alarm and nurse

call function if nurse call cable is plugged into the alarm and wall jack. Activate the alarm

(remove pressure from sensor, unfasten chair belt sensor, and make sure the nurse call light

for the proper bed and room activate in the appropriate nurse’s station location. Remove

the cable from the wall jack and make sure the visual or audible alert at the nurse’s station

immediately activates.

6. Inspect sensor cord and nurse call cable (if in use) to ensure they are out of the

footpath and DO NOT pose a tripping hazard.

Fig. 30 - Neck of Sensor Pad

NECK

SENSOR

CABLE