Choosing the Right Method for Patient Monitoring

The Sitter Elite allows you to monitor a patient in sensor mode or optional cord & magnet

mode.

NOTE: Both methods CANNOT be used at the same time. Select the mode best

suited to patient needs and risk assessment. This will vary for each patient based on:

• Activity level;

• Dexterity;

• Nighttime toileting habits;

• Cognition;

And other possible factors. Consult a qualified medical authority if you have questions about

the right option for your patient.

The Posey Sitter Elite is an electronic device. It may fail to work if subjected to

severe shock, such as being dropped, or immersed in liquid. To reduce the risk of serious injury

or death, test the alarm and sensor for proper operation prior to putting in service with a

patient, and each time before leaving the patient unattended. If the alarm and/or sensor

do not function properly, remove the alarm and sensor from service and replace them

with a properly functioning alarm and/or sensor. DO NOT use the alarm, sensor or magnet

if it does not activate each time weight is removed from the sensor, the chair belt sensor is

unfastened, or magnet is removed from face plate (see pages 22-24).

Monitoring with a Sensor

Magnet and Sensor Modes CANNOT be used at the same time. A connected

sensor will ALWAYS override the magnet. If you are using a sensor cord for monitoring

with the alarm, ALWAYS remove the magnet and store it in a safe place while the sensor is

being used.
If you are monitoring with the sensor feature and the magnet is attached, the alarm will

not activate as a failsafe if the sensor cord is removed.
The following instructions will help you set up and safely use the Sitter Elite with a sensor.
TIPS TO PROTECT SENSORS FROM DAMAGE

To avoid inconvenience to staff and patients, and to protect sensors from damage, you

should follow these steps:
• Only use Posey sensors with the Posey alarm.

• When routing sensor cord to alarm, check that there is no stress

on cord. Cord must be clear of all moving parts of bed or chair to

prevent sensor failure.

NEVER jerk or pull on the cord to remove RJ11 plug. Doing so will

damage cord wires or plug.

ALWAYS use the plastic tab to release plug (fig. 28).

FAILSAFE FEATURE

The Posey Sitter Elite contains a “failsafe” feature that activates the Posey alarm if the

sensor is removed from the alarm when the power is on. This “failsafe” will not work if the

magnet is attached to the magnet plate.