Sitter, The suspend button (sensor use only), 16 sitter elite – Posey Sitter Elite® User Manual

Page 16

Advertising
background image



16

Sitter Elite

The SUSPEND Button (Sensor Use ONLY)

NOTE: SUSPEND feature will not work unless sensor is plugged into the alarm.
The SUSPEND feature is indicated when extended patient care is required. This feature allows

you additional time to perform that care without activating the alarm.

To Place the Alarm on SUSPEND:

1. Press HOLD/SUSPEND button on the top of alarm (fig. 18) for

3 seconds. The HOLD/SUSPEND LED will begin flashing RED

at

½-

second intervals if the alarm is in SUSPEND mode (fig.

18a). The green LED indicating mode will continue to flash,

but the alarm is not monitoring.

2. You have 5 minutes to assist patient into or out of bed or

chair before alarm returns to monitoring mode.

3. After 5 minutes:

• If weight is present on the sensor, or chair belt sensor is

connected you will hear a single “beep” and monitoring

will begin.

ALWAYS verify the green light is flashing, and

the alarm and sensors are monitoring before leaving patient

unattended.

• If there is no weight on sensor, or chair belt sensor is not

connected, alarm will remain on SUSPEND.

4. The SUSPEND feature:

• Allows patient to be away from bed or chair for extended

periods without alarm activating (e.g., for meals, therapy,

toileting etc.). When patient returns and weight is applied to

sensor, or chair belt sensor is connected, alarm will “beep”

once to indicate monitoring has resumed. The red LED below

HOLD/SUSPEND will no longer be flashing.

• Helps ensure continuity of care when there are several

caregivers.

5. To take alarm out of SUSPEND:

• Apply weight to sensor, or connect chair belt sensor, or

• Press SUSPEND button once.

Alarm will “beep” once to indicate monitoring has resumed.
Flashing red HOLD/SUSPEND LED will no longer be flashing. The green light indicating

mode will blink, indicating the alarm is in monitoring mode.

Test the alarm and sensor for proper operation prior to putting in service with

a patient, and each time before leaving the patient unattended. If the alarm and/or sensor

do not function properly, remove the alarm and sensor from service and replace them with a

properly functioning alarm and/or sensor (see pages 22-24). Ensure the green LED, indicating

MODE, is blinking and the red HOLD/SUSPEND LED is no longer flashing red.

VOICE

&

TONE

TONE

MUTE

VOICE

NURSE C

ALL

MIC

HOLD

&

SUSPEND

HOLD/SUSP

END

HOLD/SUSPEND Button

Fig. 18

H O L D / S U S P E N D

®

Sitter

HOLD/SUSPEND

Indicator LED

Fig. 18a