List of serial commands, Table 6, Basic serial commands – Vaisala HMM100 User Manual

Page 32

Advertising
background image

User's Guide _______________________________________________________________________

30 ___________________________________________________________________ M211024EN-C

List of Serial Commands

All commands can be issued either in uppercase or lowercase.

The notation <cr> refers to pressing the carriage return (ENTER) key on

your computer keyboard. Enter a <cr> to clear the command buffer

before starting to enter commands.

Table 6

Basic Serial Commands

Command

Description

?

Show module information

??

Show module information (when in poll

mode)

ADDR

[0 ... 99]

Set the module address (for POLL mode)

CALCS

Show measured quantities

CLOSE

Close the temporary connection (returns to

POLL mode)

ECHO [ON/OFF]

Set terminal echo

ENV

Set environmental parameters

ERRS

Display active errors

HELP

List available commands

INTV [

0 ... 255 S/MIN/H]

Set the continuous output interval

(for RUN mode)

OPEN

[0 ... 99]

Open a temporary connection to a POLL

mode device

PASS [passcode]

Access advanced commands (requires

passcode)

R

Start the continuous outputting

RESET

Resets the module

RESTORE

Restores the latest saved settings

(overwrites unsaved settings)

S

Stop the continuous outputting

SAVE

Save changed settings to memory

SEND [0 ... 99]

Output the reading once

SERI [baud p d s]

User Port settings (Default: 19200 N 8 1)

baud: 300 ... 115200

STATUS

Show analog output status

SYSTEM

Show module information and build date

VERS

Show software version of the module

XPRES

Set the value for pressure compensations,

temporarily