View User Adjustment Parameters

Use the L command to view the current user adjustment parameters and

trimmer adjustment. This command is useful for checking the currently

applied correction.

L<cr>

The output values are as follows:

- Cp offset and gain: capacitance, calibrated using the CRH command
- RH trimmer offset and gain: correction by the trimmers
- RH offset and gain: calibrated using the CRH command
- T offset and gain: calibrated using the CT command
Example (shows default values, no user calibration done):

>l

Cp offset : 0

Cp gain : 1

RH offset trimmer : 0

RH gain trimmer : 1

RH offset : 0

RH gain : 1

T offset : 0

T gain : 1

>

Set User Adjustment Parameters

Use the LI command to set the user adjustment parameters directly. This

command is useful for restoring the desired adjustments, and removing

the effects of calibration without having to perform the factory reset.

LI<cr>

After entering the command, you will be prompted to enter new values

one at a time. Press ENTER without giving a value to move to the next

parameter.

NOTE

Remember to save the changes!