Figure 14, Tightening the clasp nut, Installing the hmp368 – Vaisala HMT360N User Manual

Page 34

Advertising
background image

User's Guide ______________________________________________________________________

32 __________________________________________________________________ M210744EN-C

0505-276

Figure 14

Tightening the Clasp Nut

Installing the HMP368 Probe Through a Ball Valve
Assembly

The ball valve installation kit (Vaisala order code: BALLVALVE-1) is
preferred when connecting the probe to a pressurized process or
pipeline. Use the ball valve set or a 1/2" ball valve assembly with a ball
hole of ø14 mm or more. If you install the probe (ø12 mm) in a process
pipe, please note that the nominal size of the pipe must be at least 1 inch
(2.54 cm). Use the manual press handle to press the probe into the
pressurized (< 10 bar) process or pipeline.

The following numbers refer to

Figure 14 on page 32

:

1

=

Probe

2

=

Clasp nut

3

=

Fitting screw

4

=

Pen

NOTE

Take care not to tighten the clasp nut more than 60° to avoid
difficulties when opening it.