Chapter 6 __________________________________________________ Calibration and Adjustment

VAISALA _______________________________________________________________________ 63

CHAPTER 6

CALIBRATION AND ADJUSTMENT

This chapter contains instructions for checking the calibration and
adjusting this product.

In this User's Guide the term "calibration" refers to comparing the
device's reading against a reference concentration. "Adjustment" refers
to changing the device's reading to correspond to the reference
concentration.

Calibration Interval

HMT360 is calibrated as shipped from the factory. Typical calibration
interval is one year. In demanding applications it may be advisable to
make the first calibration check earlier.

Factory Calibration and Adjustment

The device (or the probe only) can be sent to Vaisala Service Centers
for calibration and adjustment. See section

Technical Support on page

74

for contact information.

NOTE

HMT360 for natural gas must always be adjusted in a Vaisala Service
Center.