S scaling analog outputs – Vaisala HMT360N User Manual

Page 56

Advertising
background image

User's Guide ______________________________________________________________________

54 __________________________________________________________________ M210744EN-C

Turn the internal DIP switch Calibration Enabled/Disabled to
position ON before the selecting the analog outputs and return it to
position OFF after making the selection. Output quantities and their
abbreviations are listed in the table below:

Example:

>asel td t<cr>

Ch1 Td lo: -40.00 ’C

Ch1 Td hi: 100.00 ’C

Ch2 T lo: -40.00 ’C

Ch2 T hi: 100.00 ’C

>

S Scaling Analog Outputs

Syntax: Szz aa.a bb.b<cr>

Turn the internal DIP switch Calibration Enabled/Disabled to
position ON before the scaling the analog outputs and return it to
position OFF after scaling.

Quantity

Abbreviation

temperature

T

dewpoint temperature

Td

volume concentration

ppmv

water content

w (when scaling water content, the
abbreviation is "ng1", see note
below)

NOTE

When scaling water content w (mg/m

3

or lb/mmscf) with the natural

gas transmitter, issue command SNG1.

where

zz =

Quantity (T, Td, ppm, w)

aa.a =

Lower limit of the quantity

bb.b=

Upper limit of the quantity