ADDING OPTION CARDS

The CUB5T meters can be fitted with optional relay card and/or serial

communications cards. The details for the plug-in cards can be reviewed in the
specification section below. The plug-in cards, that are sold separately,

can be

installed initially or at a later date.

RELAY CARD

Type: Single FORM-C relay
Isolation To Sensor & User Input Commons: 1400 Vrms for 1 min.

Working Voltage: 150 Vrms

Contact Rating: 1 amp @ 30 VDC resistive; 0.3 amp @ 125 VAC resistive
Life Expectancy: 100,000 minimum operations
Response Time:

Turn On Time: 4 msec max.
Turn Off Time: 4 msec max.

Time Accuracy: ± 0.01%

WARNING: Disconnect all power to the meter before

installing Plug-in card.

RS485 SERIAL COMMUNICATIONS CARD

Type: RS485 multi-point balanced interface (non-isolated)
Baud Rate: 300 to 38400
Data Format: 7/8 bits; odd, even, or no parity
Bus Address: 0 to 99; max 32 meters per line
Transmit Delay: Selectable. 2 msec min. or 50 msec min.

RS232 SERIAL COMMUNICATIONS CARD

Type: RS232 half duplex (non-isolated)
Baud Rate: 300 to 38400
Data Format: 7/8 bits; odd, even, or no parity

O

PTIONAL

P

LUG

-

IN

C

ARDS

1.0 I

NSTALLING

THE

M

ETER

INSTALLATION

The meter meets NEMA 4X/IP65 requirements when properly installed. The

unit is intended to be mounted into an enclosed panel. Prepare the panel cutout
to the dimensions shown. Remove the panel latch from the unit. Slide the panel
gasket over the rear of the unit to the back of the bezel. The unit should be
installed fully assembled. Insert the unit into the panel cutout.

While holding the unit in place, push the panel latch over the rear of the unit

so that the tabs of the panel latch engage in the slots on the case. The panel latch
should be engaged in the farthest forward slot possible. To achieve a proper seal,
tighten the latch screws evenly until the unit is snug in the panel (Torque to
approx. 28 to 36 in-oz [0.202 to 0.26 N-m]). Do not over-tighten the screws.

INSTALLATION ENVIRONMENT

The unit should be installed in a location that does not exceed the operating

temperature and provides good air circulation. Placing the unit near devices that
generate excessive heat should be avoided.

The bezel should only be cleaned with a soft cloth and neutral soap product.

Do NOT use solvents. Continuous exposure to direct sunlight may accelerate the
aging process of the bezel.

Do not use tools of any kind (screwdrivers, pens, pencils, etc.) to operate the

keypad of the unit.

2.0 DIP S

WITCHES

2.68

(68 )

+.8

1.30

-.0

(33 )

-.000

+.6

-.0

-.000

+.024

+.025

The DIP switches on the main circuit board are not used with the CUB5T and

must be left in the factory set position (all down). Setting any switch to the up
position may cause improper operation of the meter.