SPECIFICATIONS

FLOW MEDIUM: Please note that GFC Mass Flow Controllers are designed to work with
clean gases only. Never try to meter or control flow rates of liquids.

CALIBRATIONS: Supplied at Standard Conditions (14.7 psia and 70

F

F), or Normal

Conditions (0

F

C and 1.01 bar abs) unless otherwise requested or stated.

ENVIRONMENTAL (per IEC 664): Installation Level II; Pollution Degree II.

ACCURACY:

REPEATABILITY: ±0.5% of full scale.

TEMPERATURE COEFFICIENT: 0.15% of full scale/

F

C.

PRESSURE COEFFICIENT: 0.01% of full scale/psi (0.07 bar).

RESPONSE TIME: GFC17: 300ms time constant; approximately 1 second to within

±2% of set flow rate for 25% to 100% of full scale flow.

GFC 37/47: 600ms time constant; approximately 2 seconds to within
±2% of set flow rate for 25% to 100% of full scale flow.

GFC 57/67/77: 1800ms time constant; approximately 5 seconds to
within ± 2% of set flow rate for 25% to 100% of full scale flow.

MAX GAS PRESSURE: 1000 psig (69 bars) GFC 17, 37 and 47; 500 psig (34.5 bars) GFC
57, 67 and 77. Optimum pressure is 20 psig (1.4 bars).

TURNDOWN RATIO: 40:1.

MAX DIFFERENTIAL PRESSURE: 50 psid (345 kPa).

GAS TEMPERATURE: 32

F

F to 122

F

F (0

F

C to 50

F

C).

AMBIENT TEMPERATURE:14

F

F to 122

F

F (-10

F

C to 50

F

C).

GAS RELATIVE HUMIDITY: Up to 70%.

MAXIMUM INTERNAL LEAK: 0.5% FS.

ACCURACY % FS

OPTIONAL ENHANCED ACCURACY % FS

MODEL

GFC
17, 37

GFC
47, 57, 67, 77

MODEL

GFC 17

GFC
37, 47, 57, 67, 77

FLOW
RANGE

2.5-100% 20-100% 0-20%

FLOW
RANGE

2.5-100% 20-100% 0-20%

ACCURACY ±1.5%

±1.5%

±3%

ACCURACY

±1%

±1%

REF DATA
with ±1%