Aalborg GFC User Manual

Page 13

Advertising
background image

GFC 17/37/47 CAUTION

CAUTION: If the valve is left in the AUTO (control) or OPEN (PURGE)
mode for an extended period of time, it may become warm or even hot
to the touch. Use care in avoiding direct contact with the valve during
operation.

Do not run GFC 17/37/47 models for extended periods of time with the valve in
AUTO or PURGE mode without the flow of gas through the transducer. Doing so
may result in up to 2% f.s. shift in calibration.

ƽ

9

5.

OPERATING INSTRUCTIONS

5.1

Preparation and Warm Up

It is assumed that the Mass Flow Controller has been correctly installed and thor-
oughly leak tested as described in section (2). Shut the flow source OFF. Apply
power to the unit via the 15-pin “D” connector. Use a power supply that is between
+12 and +15 VDC with at least 800 mA current capacity (or optionally, for models
GFC17/37/47 only, +24 VDC 650 mA). Allow the Mass Flow Controller to warm-
up for at least 15 minutes.

During initial powering of the GFC transducer, the flow output signal will be indi-
cating a higher than usual output. This is indication that the GFC transducer has
not yet attained its minimum operating temperature. This condition will automat-
ically cancel within a few minutes and the transducer should eventually zero.

If after the 15 minutes warm-up period, the display still indicates a reading of less
than ± 3.0 % of F.S., readjust the ZERO potentiometer [R34] through the access
window. Before zero adjustment temporarily disconnect the gas source, to ensure
that no seepage or leak occurs in to the meter.

TABLE IV PRESSURE DROPS

MODEL

FLOW RATE

[liters/min]

MAXIMUM PRESSURE DROP

[mm H2O]

[psid]

[mbar]

GFC 17

UP to 10

720

1.06

75

GFC 37

15

2630

3.87

266

20

1360

2.00

138

30

2380

3.50

241

40

3740

5.50

379

50

5440

8.00

551

GFC 47

60

7480

11.00

758

100

12850

18.89

1302

GFC 57

200

7031

10.00

690

GFC 67

500

8437

12.00

827

GFC 77

1000

10547

15.00

1034