INDICATION

LIKELY REASON

REMEDY

calibration off

gas metered is not the same as

use matched calibration

what meter was calibrated for

composition of gas changed

see K factor tables in APPENDIX 2

gas leak

locate and correct

pc board defective

return to factory for replacement

RFE dirty

flush clean or disassemble to
remove impediments

occluded sensor tube

flush clean or disassemble to
remove impediments or return to
factory for replacement

filter screen obstructed

flush clean or disassemble to

at inlet

remove impediments or replace

transducer is not

check for any tilt or change in the

mounted properly

mounting of the transducer;
generally, units are calibrated for
horizontal installation
(relative to the sensor tube)

GFC valve does

incorrect valve adjustment

re-adjust valve (section 7.2.4)

not work
in open position

pc board defect

return to factory for replacement

cable or connectors

check cable and connectors

malfunction

or replace

differential pressure too high

decrease pressure to correct level

insufficient inlet pressure

adjust appropriately

GFC valve does incorrect valve adjustment

re-adjust valve (section 7.2.4)

not work in
closed position

pc board defect

return to factory for replacement

cable or connectors

check cable and connectors

malfunction

or replace

orifice obstructed

disassemble to remove
impediments or return to factory