22

8.2

Troubleshooting Guide

INDICATION

LIKELY REASON

REMEDY

lack of reading

power supply off

check connection of power supply

or output

fuse blown

disconnect transducer from
power supply; remove the
shorting condition or check
polarities; fuse resets
automatically

filter screen

flush clean or disassemble to

obstructed at inlet

remove impediments or replace

occluded sensor tube

flush clean or disassemble to
remove impediments or return to
factory for replacement

pc board defect

return to factory for replacement

valve adjustment wrong

re-adjust valve (section 7.2.4)

flow reading

inadequate gas pressure

apply appropriate gas pressure

does not
coincide with

filter screen obstructed

flush clean or disassemble to

the setpoint

at inlet

remove impediments or replace

ground loop

signal and power supply
commons are different

no response

inadequate gas pressure

apply appropriate gas pressure

to setpoint

cable or connector malfunction

check cables and all connections
or replace

setpoint is too low

re adjust setpoint or disable 2%

(<2% of full scale)

cutoff feature (section 2.2)

valve adjustment wrong

re-adjust valve (section 7.2.4)

unstable or

gas leak

locate and correct

no zero reading

pc board defective

return to factory for replacement

full scale output defective sensor

return to factory for replacement

at “no flow”
condition or

gas leak

locate and repair

with valve
closed