3 troubleshooting guide – Aalborg DFC Digital User Manual

Page 21

Advertising
background image

8.3

Troubleshooting Guide

18

Remedy

check connection of power supply

disconnect DFC transducer from power
supply; remove the shorting condition
or check polarities;
fuse resets automatically

disconnect power cord from AC
supply; remove and inspect fuses at
AC power input connector of SDPROC;
replace as necessary

REMOVE CAUSE OF SHORT CIRCUIT!

flush clean or disassemble to remove
impediments or replace

flush clean or disassemble to remove
impediments or return to factory for
replacement

return to factory for replacement

re-adjust valve (section 6.2.4)

disconnect DFC transducer from
power supply; remove the shorting
condition or check polarities; fuse
resets automatically

REMOVE CAUSE OF SHORT CIRCUIT!

apply appropriate gas pressure

flush clean or disassemble to remove
impediments or replace

signal and power supply commons are
different

apply appropriate gas pressure

check cables and all connections or
replace

re-adjust set point

re-adjust valve (section 6.2.4)

locate and correct

return to factory for replacement

Likely Reason

power supply off

fuse blown
(DFC)

fuse blown
(SDPROC)

filter screen obstructed at inlet

occluded sensor tube

pc board defect

valve adjustment wrong

fuse blown
(DFC)

inadequate gas pressure

filter screen obstructed at inlet

ground loop

inadequate gas pressure

cable or connector
malfunction

set point is too low
(<2% of full scale)

valve adjustment wrong

gas leak

pc board defective

Indication

lack of
reading
or output

output reads
at (+) or (-)
saturation
only

flow reading
does not
coincide with
the set point
(DFC models
only)

no response
to set point
(DFC models
only)

unstable or no
zero reading