Aalborg DFC Digital User Manual

Page 22

Advertising
background image

19

Indication

full scale
output at
"no flow"
condition or
with valve
closed

calibration off

DFC valve
does not work
in open
position

DFC valve
does not
work in close
position

Remedy

return to factory for replacement

locate and repair

use matched calibration

see K factor tables in APPENDIX 2

locate and correct

return to factory for replacement

flush clean or disassemble to remove
impediments

flush clean or disassemble to remove
impediments or return to factory for
replacement

flush clean or disassemble to remove
impediments or replace

check for any tilt or change in the
mounting of the transducer; generally,
units are calibrated for horizontal
installation (relative to the sensor tube)

re-adjust valve (section 6.2.4)

return to factory for replacement

check cable and connectors or replace

decrease pressure to correct level

adjust appropriately

re-adjust valve (section 6.2.4)

return to factory for replacement

check cable and connectors or replace

disassemble to remove impediments
or return to factory

Likely Reason

defective sensor

gas Leak

gas metered is not the same
as what meter was calibrated
for

composition of gas changed

gas leak

pc board defective

RFE dirty

occluded sensor tube

filter screen obstructed at inlet

transducer is not mounted
properly

incorrect valve adjustment

pc board defect

cable or connectors
malfunction

differential pressure too high

insufficient inlet pressure

incorrect valve adjustment

pc board defect

cable or connectors
malfunction

orifice obstructed