Aalborg DFC Digital User Manual

Page 8

Advertising
background image

PIN FUNCTION

1

+15 VDC Power Supply

2

0-5 VDC Flow Signal (4-20mA Option)

3

0-5 VDC Set Point Input (4-20mA Option)

4

Force Valve Open Control

5

Force Valve Closed Control

6 (Reserved)
7 (Reserved)
8

Relay No. 1 - Common Contact

9

Relay No. 1 - Normally Open Contact

10

Relay No. 2 - Normally Closed Contact

11

RS485 (-) (Optional RS232 TX)

12 (No

Connection)

13 Chassis

Ground

14

-15 VDC Power Supply

15

Common, Signal Ground For Pin 2

16

Common, Signal Ground For Pin 3

17

(Optional) RS232 Common

18

Common, Power Supply

19 Common
20 Common
21

Relay No. 1 - Normally Closed Contact

22

Relay No. 2 - Common Contact

23

Relay No. 2 - Normally Open Contact

24

RS485 (+) (Optional RS232 RX)

25

Return for Pin 2 (Optional 4-20 mA Only)

FIGURE b-2, DFC 25 PIN "D" CONNECTOR CONFIGURATION

Important notes:

In general, "D" Connector numbering patterns are standardized. There are, how-
ever, some connectors with nonconforming patterns and the numbering sequence
on your mating connector may or may not coincide with the numbering sequence
shown in our pin configuration table above. It is imperative that you match the
appropriate wires in accordance with the correct sequence regardless of the par-
ticular numbers displayed on your mating connector.

Make sure power is OFF when connecting or disconnecting any cables in the system.

The (+) and (-) power inputs are each protected by a 500mA M (medium time-lag)
resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut
power to the flow transducer circuit. Disconnect the power to the unit, remove the
faulty condition, and reconnect the power. The fuse will reset once the faulty con-
dition has been removed.

Cable length may not exceed 9.5 feet (3 meters).
Use of the DFC flow transducer in a manner other than that specified in this manu-
al or in writing from Aalborg

7, may impair the protection provided by the equipment.