5 troubleshoot, Troubleshoot – Campbell Scientific AVW200-series Vibrating Wire Interfaces User Manual
Page 41

AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules
5.5 Troubleshoot
The Troubleshoot tool in DevConfig can be used to evaluate the frequency
spectrum of a sensor and to determine the most appropriate beginning and
ending frequencies for a sensor.
To access the Troubleshoot Tool, use the steps listed below:
1. Follow the connection procedure provided in Section 5.1, Connecting to
DevConfig.
2. Click the Troubleshoot tab at the top of the DevConfig opening window
FIGURE 5-7. Opening Page of the Troubleshoot Tool
3. Click the Options tab at the bottom of the Troubleshoot screen to set the
begin and end frequencies and the excitation voltage you wish to test for a
given sensor (FIGURE 5-8). You may also choose to poll (default) or not
to poll the time series data from this Options window by checking or
unchecking the poll time series box.