2 extended sdi-12 commands, 3 use with multiplexers, Extended sdi-12 commands – Campbell Scientific AVW200-series Vibrating Wire Interfaces User Manual

Page 52: Use with multiplexers

Advertising
background image

AVW200-series 2-Channel Vibrating Wire Spectrum Analyzer Modules

If a check summed command fails, a NAN will be returned and the command

will be retried.

Mult, Offset

The Mult and Offset parameters are each a constant,

variable, array, or expression by which to scale the results of

the measurement.

6.2.2 Extended SDI-12 Commands

Extended SDI-12 commands can be used to change the beginning frequency,

ending frequency, and excitation voltage of the vibrating wire sensors attached

to the AVW200. However, these extended SDI-12 commands only work for

the next measurement command. By default, standard SDI-12 measurement

commands use the begin/end/excite voltage settings specified in the AVW200

settings using DevConfig (see Section 4, Connections). However, after issuing

an extended SDI-12 command, the next measurement will use the bbbb,eeee

and vvvv values specified in the extended command.

An example of an extended command is: 0XVW450,5000,1! This command

configures the next measurement with Begin Freq=450, End Freq = 5000 and

5Volt excitation. For example, an “aM!” command that immediately follows

an extended command will use the settings specified in the extended command.

The second and remaining measurements will revert back to the settings

specified via DevConfig.

The SDI-12 aI! command is used to obtain information about a specific sensor.

When executed against the AVW200-series device, the following information

is returned:

1. The SDI-12 address of the device (1 character)

2. The SDI-12 implementation version (2 characters, "13" is interpreted as 1.3)

3. The Vendor ID "Campbell" (8 Characters)

4. The Sensor Model "AVW2xx" (6 characters)

5. Sensor Version "000" (3 characters)

6. The Serial Number (up to 13 characters), which is the version of the

currently loaded firmware followed by an equals sign and the device's serial

number (“Std.04=875321")

6.2.3 Use with Multiplexers

The AVW200 interface module cannot control multiplexers in the SDI-12

communication mode. Hence, when communicating with the AVW200 via

SDI-12, multiplexers attached to the AVW200 must be controlled by the

datalogger. This is achieved by using PortSet instructions in the datalogger

program and by connecting the clock and reset lines of the multilplexers to

control ports on the datalogger. When using SDI-12 with the AVW200, the

clock and reset lines on the AVW200 are not used.

Read more! Example programs of SDI-12 are provided in Section 7.5.

44